COULTER® LH 750 ANALYZER
Report
- Report Number
- 1061932-2013-00822
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 19, 2019
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE CONFIRMED THAT THE INSTRUMENT WAS GENERATING HIGH HGB RESULTS. PER PHONE CONVERSATION, THE FSE STATED THAT HE REPLACED SLUGGISH ACTUATORS ON VL12C (BATH DRAINS) AND VL57A (PROBE/NEEDLE DRAIN VACUUM) TO RESOLVE THE ISSUE. THE FSE ALSO REPLACED THE HGB CUVETTE AND HGB LAMP AS PREVENTATIVE MAINTENANCE. THE INSTRUMENT WAS VERIFIED AND TESTED WITHIN SPECIFICATIONS. THE FAILURE MODE OF THE ERRONEOUS HGB AND MCHC RESULTS WAS RELATED TO SLUGGISH ACTUATORS ON VL57A (PROBE/NEEDLE DRAIN VACUUM) WHICH WAS NOT ALLOWING THE PROPER DRAINAGE OF VL12C (BATH DRAINS) RESULTING IN CARRY OVER. THIS EVENT IS ALSO A USE ERROR; NORMAL LEVEL CONTROL SAMPLES WERE RUN BUT FAILED IMMEDIATELY BEFORE THE EVENT AND AFTER PATIENT SAMPLES WERE RUN. PER LABELING, THE LH750 INSTRUCTIONS FOR USE (IFU) STATES FAILURE TO RECOVER CONTROL VALUES WITHIN THE EXPECTED RANGE IN EITHER MODE SHOULD BE INVESTIGATED BEFORE PROCEEDING WITH THE ANALYSIS OF PATIENT SAMPLES.
THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THEY OBTAINED ERRATIC HEMOGLOBIN (HGB) RESULTS ON TWO PATIENT SAMPLES TESTED ON THE COULTER LH 750 ANALYZER. DATA PROVIDED BY THE CUSTOMER FOR REVIEW INDICATED HEMATOCRIT/HEMOGLOBIN (H/H) CHECK FAILURES WITH LOWER HGB AND MEAN CORPUSCULAR HEMOGLOBIN CONCENTRATION (MCHC) RESULTS ON THE INITIAL RUN COMPARED TO THE RESULTS OBTAINED ON THE RE-RUN WITH NO INSTRUMENT GENERATED MESSAGES. THE 5C CELL CONTROLS WERE RUN IMMEDIATELY BEFORE AND AFTER PATIENT SAMPLES A AND B WERE RUN, WHICH DEMONSTRATED HGB, MEAN CORPUSCULAR HEMOGLOBIN (MCH) AND MCHC FAILURE FOR THE NORMAL CONTROL. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT THE LABORATORY, AND THERE WAS NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. THE RE-RUNS WERE CONSIDERED CORRECT FOR BOTH PATIENT SAMPLES. PATIENTS' RESULTS ARE PROVIDED IN THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212214 | COULTER® LH 750 ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |