FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112526 · Received May 14, 2013

Report

Report Number
2032227-2013-01936
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
May 3, 2013
Report Date
May 3, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A FROZEN DISPLAY WITH A CONSTANT TONE DUE TO MOISTURE DAMAGE ON THE LCD AND MOTHER BOARD. UNABLE TO CONFIRM THE MOTOR ERROR ALARMS DUE TO THE FROZEN DISPLAY. MOISTURE DAMAGE WAS NOTED ON THE MOTOR ASSEMBLY. NO BLANK DISPLAY WAS NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP GAVE A MOTOR ERROR ALARM FOLLOWED BY A BLANK DISPLAY. IT WAS STATED THAT THIS CONTINUED HAPPENING IN A LOOP, AND THE INSULIN PUMP WOULD NOT GET OUT OF IT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212347 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LCAS

Patients

Seq Age Sex Outcome Treatment
1