FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3112526
·
Received May 14, 2013
Report
- Report Number
- 2032227-2013-01936
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- May 3, 2013
- Report Date
- May 3, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP HAD A FROZEN DISPLAY WITH A CONSTANT TONE DUE TO MOISTURE DAMAGE ON THE LCD AND MOTHER BOARD. UNABLE TO CONFIRM THE MOTOR ERROR ALARMS DUE TO THE FROZEN DISPLAY. MOISTURE DAMAGE WAS NOTED ON THE MOTOR ASSEMBLY. NO BLANK DISPLAY WAS NOTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP GAVE A MOTOR ERROR ALARM FOLLOWED BY A BLANK DISPLAY. IT WAS STATED THAT THIS CONTINUED HAPPENING IN A LOOP, AND THE INSULIN PUMP WOULD NOT GET OUT OF IT. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212347 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LCAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |