FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3112513
·
Received May 14, 2013
Report
- Report Number
- 2032227-2013-01944
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 19, 2013
- Report Date
- April 19, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER CALLED TO REPORT AN ALARM DURING THE MANUAL PRIME. THE CUSTOMER ALSO REPORTED THAT THE DRIVE SUPPORT CAP WAS PROTRUDING. THE REPORTED BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 126 MG/DL. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213048 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |