FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112513 · Received May 14, 2013

Report

Report Number
2032227-2013-01944
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 19, 2013
Report Date
April 19, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT AN ALARM DURING THE MANUAL PRIME. THE CUSTOMER ALSO REPORTED THAT THE DRIVE SUPPORT CAP WAS PROTRUDING. THE REPORTED BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 126 MG/DL. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213048 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAH

Patients

Seq Age Sex Outcome Treatment
1 38 YR