PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3112511 · Received May 14, 2013

Report

Report Number
2032227-2013-01937
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE PRIME TEST DUE TO A LOOSE/FLUSH MOTOR SUPPORT DISK. MOISTURE DAMAGE WAS NOTED AT THE MOTOR ASSEMBLY. THE INSULIN PUMP HAD MISSING SEGMENTS DUE TO A CRACKED ZEBRA CONNECTOR AT THE LCD BOARD. THE INSULIN PUMP WAS RECEIVED WITH A SCRATCHED LCD WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN ALARM DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.