FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 3112235 · Received May 14, 2013

Report

Report Number
0001811755-2013-01110
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 18, 2013
Report Date
April 18, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE MOTOR.

Description of Event or Problem · 1

THE MICRO SAGITTAL SAW WAS TESTED DURING ROUTINE SERVICING. UPON EVALUATION IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212868 MICRO SAGITTAL SAW DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO