LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01047
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- November 21, 2013
- Report Date
- April 11, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADJUST BELT OR CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED THE FALL-OFF TEST. THE WHITE DRIVEN GROUND WIRE WAS OPEN IN THE TRUNK CABLE INSULATION. THE CAUSE FOR THE ALARMS WAS THE OPEN WIRE. THE ROOT CAUSE FOR THE OPEN WIRE COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM EXCESSIVE FORCE ON THE ELECTRODE BELT TRUNK CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) DISTRIBUTOR CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT AN UNKNOWN (B)(6) PATIENT WAS RECEIVING CHECK ELECTRODE BELT ALARMS. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193541 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |