LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01199
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- April 3, 2013
- Report Date
- May 1, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY FAULTS, CHARGER FAULTS) WAS CONFIRMED. UPON EVALUATION, THERE WAS CONTAMINATION ON THE BATTERY BOARD INSIDE THE CHARGER/MODEM. THE CAUSE OF THE INABILITY TO CHARGE A BATTERY IS THE CONTAMINATION ON THE BOARD INSIDE THE CHARGER/MODEM. THE ROOT CAUSE OF THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY LIQUID INGRESS OF AN UNKNOWN CONTAMINANT. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT CHARGER/MODEM.
A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER CHARGER/MODEM WAS DISPLAYING "BATTERY HAS A PROBLEM". THE PATIENT WAS ISSUED A REPLACEMENT CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 195529 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |