FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3112149 · Received May 3, 2013

Report

Report Number
3008642652-2013-01193
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 14, 2013
Report Date
May 1, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. AS RECEIVED, THE TRUNK CABLE CONNECTOR WAS CRACKED DAMAGING INTERNAL WIRES. UPON EVALUATION, THE GREEN (3.3V) WIRE WAS OPEN IN THE TRUNK CABLE CONNECTOR. THE CAUSE OF THE SERVICE CODE 204 IS A DISRUPTION IN COMMUNICATION BETWEEN THE MONITOR AND ELECTRODE BELT. THE CAUSE OF THE DISRUPTION IN COMMUNICATION IS THE OPEN GREEN WIRE. THE CAUSE OF THE OPEN WIRE IS THE DAMAGED TRUNK CABLE CONNECTOR. THE ROOT CAUSE OF THE DAMAGED TRUNK CABLE CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT CONNECTOR. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT'S WIFE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A SERVICE CODE 204. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195528 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR