FDA Adverse Event Malfunction Summary report: N

8CM HEAVY DUTY MEDIUM ATTACHMENT

MDR report key: 3111883 · Received May 7, 2013

Report

Report Number
1045834-2013-02090
Event Type
Malfunction
Date Received
May 7, 2013
Date of Event
April 5, 2013
Report Date
April 8, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ATTACHMENT DEVICE WAS REC'D FOR EVAL. THE ATTACHMENT DEVICE WAS TESTED AND IT WAS OBSERVED THAT THE ATTACHMENT DEVICE FAILED THE CUTTER INSERTION AND LOCK OPERATIONS TEST. EVIDENCE INDICATED THIS WAS DUE TO USAGE WEAR OVER TIME. THE REPORTED CONDITION WAS CONFIRMED. IF ADD'L INFO IS REC'D, A SUPPLEMENTAL REPORT WILL BE SENT.

Description of Event or Problem · 1

REPORT REC'D FROM THE USA STATING THAT IT WAS "HARD TO ATTACH THE DRILL AND HARD TO INSERT THE BIT" TO THE ATTACHMENT DEVICE. THE ALLEGED MALFUNCTION WAS DISCOVERED DURING SET-UP FOR A CRANIOTOMY PROCEDURE BEFORE THE PT ENTERED THE ROOM. THERE WAS A 1-2 MIN DELAY IN SURGERY. AN IDENTICAL ATTACHMENT DEVICE WAS AVAILABLE. THERE WAS NO ADD'L INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200003 8CM HEAVY DUTY MEDIUM ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1