FDA Adverse Event
Malfunction
Summary report: N
8CM HEAVY DUTY MEDIUM ATTACHMENT
MDR report key: 3111883
·
Received May 7, 2013
Report
- Report Number
- 1045834-2013-02090
- Event Type
- Malfunction
- Date Received
- May 7, 2013
- Date of Event
- April 5, 2013
- Report Date
- April 8, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ATTACHMENT DEVICE WAS REC'D FOR EVAL. THE ATTACHMENT DEVICE WAS TESTED AND IT WAS OBSERVED THAT THE ATTACHMENT DEVICE FAILED THE CUTTER INSERTION AND LOCK OPERATIONS TEST. EVIDENCE INDICATED THIS WAS DUE TO USAGE WEAR OVER TIME. THE REPORTED CONDITION WAS CONFIRMED. IF ADD'L INFO IS REC'D, A SUPPLEMENTAL REPORT WILL BE SENT.
Description of Event or Problem · 1
REPORT REC'D FROM THE USA STATING THAT IT WAS "HARD TO ATTACH THE DRILL AND HARD TO INSERT THE BIT" TO THE ATTACHMENT DEVICE. THE ALLEGED MALFUNCTION WAS DISCOVERED DURING SET-UP FOR A CRANIOTOMY PROCEDURE BEFORE THE PT ENTERED THE ROOM. THERE WAS A 1-2 MIN DELAY IN SURGERY. AN IDENTICAL ATTACHMENT DEVICE WAS AVAILABLE. THERE WAS NO ADD'L INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200003 | 8CM HEAVY DUTY MEDIUM ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |