FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 3111872 · Received May 7, 2013

Report

Report Number
1828100-2013-00454
Event Type
Malfunction
Date Received
May 7, 2013
Date of Event
April 16, 2013
Report Date
April 16, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS RELATED TO MDR #1828100-2013-00351. THE SUBSIDIARY FIELD SERVICE REP (FSR) WENT TO THE HOSPITAL TO REPLACE THE ELBOW. THE FSR TESTED THE ICE COMPRESSOR AGAIN, BUT IT DID NOT START AFTER FIVE MINS, WE DID THE SAME EVAL ON (B)(4) 2013 (MDR #1828100-2013-00351). THE ICE COMPRESSOR WAS STARTED AND RAN AGAIN. THE FSR DECIDED TO CHECK THE UNIT THE NEXT DAY. UPON ARRIVAL THE FOLLOWING DAY, THE FSR FOUND A GROUP OF AIR BUBBLES FROM THE AIR COMPRESSOR IN THE LARGE TANK BUT THE ICE COMPRESSOR WOULD NOT START. WHEN THE PRESSURE SWITCH WAS PRESSED THE THIRD TIME, THE ICE COMPRESSOR FINALLY STARTED AND RAN. THE FSR THOUGHT THE 1/4 INCH OD TUBING WAS TOO LONG AND CUT AROUND 60CM OFF, BUT DID NOT SOLVE THE PROBLEM. THE FSR INSTALLED THE NEW PRESSURE SWITCH BOARD AND COMPRESSOR DELAY BOARD. IT OPERATED AGAIN, BUT WHEN WE TRIED THE THIRD TIME, THE SAME PROBLEM OCCURRED AGAIN. THE FSR ALSO CHECKED THE 1/4 INCH OD TUBING AND AIR COMPRESSOR. THE COMPONENTS SEEMED TO BE GOOD. WE BLEW AIR WITH MOUTH INTO THE 1/4 INCH TUBING, AIR BUBBLES WERE FOUND IN THE LARGE TANK. THE FSR FOUND THAT HE COULD PRESS THE PRESSURE SWITCH AGAIN AND AGAIN (LESS THAN FOUR TIMES) TO START THE COMPRESSOR.

Description of Event or Problem · 1

THE FIELD SERVICE REP (FSR) REPORTED THAT DURING FIELD SERVICE INSTALLATION OF THE DEVICE, THE COMPRESSOR WOULD NOT STARTUP AFTER FIVE MINS ON THE COOLER HEATER UNIT. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200036 TERUMO TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP 4416

Patients

Seq Age Sex Outcome Treatment
1