FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3111837 · Received May 14, 2013

Report

Report Number
1416980-2013-12282
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 13, 2013
Report Date
April 23, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 22:20:27. DURING NIGHT DRAIN CYCLE THREE, THE PATIENT'S ULTRAFILTRATION READING WAS 1251ML, INDICATING THE HOME PATIENT (HP) DRAINED 1251ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213354 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1