PLUM A+ DRIVER NEW 8
Report
- Report Number
- 9615050-2013-01071
- Event Type
- Malfunction
- Date Received
- May 7, 2013
- Date of Event
- April 12, 2013
- Report Date
- April 12, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 199958 | PLUM A+ DRIVER NEW 8 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST #20791, SN (B)(4) |