FDA Adverse Event
Malfunction
Summary report: N
PLMA DVC V11.51 1 N
MDR report key: 3111657
·
Received May 7, 2013
Report
- Report Number
- 9615050-2013-01081
- Event Type
- Malfunction
- Date Received
- May 7, 2013
- Date of Event
- April 6, 2013
- Report Date
- April 9, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE DOOR ROLLER WAS MISSING AND THE ROLLER PIN WAS BROKEN. AS INDICATED, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT THE DEVICE DOOR ROLLER PIN WAS BROKEN. IT WAS REPORTED THAT DURING SET UP PRIOR TO PT USE, THE NURSE NOTED THAT THE DEVICE DOOR ROLLER PIN WAS BROKEN. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 197761 | PLMA DVC V11.51 1 N | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | PLUM A+ SOFTWARE MODULE: LIST #12097, (B)(4) |