PLM A+PMP W/BSE MOD

MDR report key: 3111655 · Received May 7, 2013

Report

Report Number
9615050-2013-01054
Event Type
Malfunction
Date Received
May 7, 2013
Date of Event
March 1, 2013
Report Date
April 11, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K011442
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT DURING TESTING, THE DEVICE DID NOT ALARM WITH AN E321 ERROR CODE; HOWEVER, THE BATTERY WAS FOUND TO NOT HOLD A CHARGE. THE DEVICE HISTORY WAS DOWNLOADED AT THE SERVICE CENTER. A REVIEW OF THE DEVICE HISTORY DID NOT FIND ANY E321 ERROR CODES AS REPORTED BY THE CUSTOMER CONTACT. AS INDICATED, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH AN E321 (BATT CHARGER TIMEOUT) ERROR CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A NOTE THAT STATED, "DEPLETED BATTERY." THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE PUMP WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH AN E321 (BATT CHARGER TIMEOUT) ERROR CODE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
199817 PLM A+PMP W/BSE MOD 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE, LIST #12097, SN (B)(4)