FDA Adverse Event Malfunction Summary report: N

LIFECARE PCA 3 V5.06

MDR report key: 3111654 · Received May 7, 2013

Report

Report Number
9615050-2013-01053
Event Type
Malfunction
Date Received
May 7, 2013
Date of Event
January 1, 2012
Report Date
April 9, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
MEA
PMA / PMN Number
K042800
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THIS WAS DUE TO WORN THREADS ON THE HALF NUT. AS INDICATED, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. SEE ATTACHED CUSTOMER NOTIFICATION DATE 03/08/2013. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, IT WAS NOTED THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE PUMP WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197760 LIFECARE PCA 3 V5.06 80MEA MEA HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA