FDA Adverse Event
Malfunction
Summary report: N
GYNECARE X-TRACT MORCELLATOR
MDR report key: 3111628
·
Received May 14, 2013
Report
- Report Number
- 2210968-2013-05510
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- December 12, 2012
- Report Date
- April 23, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K993801
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE COUPLER WAS FOUND TO BE BENT AND LOOSE ON THE MOTOR SHAFT. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, THE DEVICE WAS MAKING A VIBRATING NOISE AND A LOUD GRINDING NOISE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212340 | GYNECARE X-TRACT MORCELLATOR | LAPAROSCOPIC MORCELLATOR | HET | ETHICON, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |