FDA Adverse Event Other Summary report: N

VISTA BASIC - IFVB

MDR report key: 3111465 · Received April 30, 2013

Report

Report Number
2523676-2013-00114
Event Type
Other
Date Received
April 30, 2013
Report Date
April 3, 2013
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
FRN
Product Problem
Yes
Report Source
Distributor report
Reporter Location
ME, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY: EVENT: BLOOD FROM DIALYSIS UNIT BACKING UP LINE DURING HEMODIALYSIS. FACILITY IS ADMINISTERING AN ANTIBIOTIC (EITHER VANCOMYCIN OR GENTAMYCIN) DURING VENOUS CHAMBER OF HEMODIALYSIS MACHINE; INFUSING VIA VISTA BASIC PUMP. THE PUMP IS SET ON HIGH OCCLUSION PRESSURE. BLOOD FROM DIALYSIS MACHINE IS BACKING UP LINE INTO BAG; NO ALARM IS SOUNDING. ALL OTHER VISA BASIC PUMPS ARE SET UP THE SAME WAY AND ARE NOT HAVING ANY INFUSION ISSUES. REF MFR REPORT: 9610825-2013-00114.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188716 VISTA BASIC - IFVB INFUSION PUMP FRN B. BRAUN MELSUNGEN AG NA NA

Patients

Seq Age Sex Outcome Treatment
1 * Other