FDA Adverse Event Injury Summary report: N

BIOHORIZON DENTAL IMPLANT DRILL

MDR report key: 3111279 · Received May 9, 2013

Report

Report Number
1060818-2013-00001
Event Type
Injury
Date Received
May 9, 2013
Date of Event
April 9, 2013
Report Date
April 11, 2013
Manufacturer
BIOHORIZONS
Product Code
DZA
PMA / PMN Number
K060267
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

DURING THE MFR'S EVAL, THE DRILL EXTENDER WAS TESTED ON AN INTERFACE VERIFICATION TEST (IVT) WITH A MATING (B)(4) SHANK. THE INVESTIGATOR CONCLUDED THE DRILL EXTENDER DID NOT WORK PROPERLY. THE FREQUENCY OF USAGE COULD NOT BE DETERMINED OR ASCERTAINED. THE DRILL EXTENDER'S (B)(4) SHANK SLIPS OUT EASILY WHEN THE DRILL EXTENDER IS POINTED DOWNWARD. THIS DRILL EXTENDER WAS PURCHASED IN (B)(4) 2011 AS PART OF A SURGICAL KIT.

Description of Event or Problem · 1

IN THE COURSE OF THE SURGICAL PROCEDURE FOR IMPLANTING AN IMPLANT, A STARTER DRILL WAS BEING USED WITH A DRILL EXTENDER (122-100). DURING THE PROCEDURE, THE CLINICIAN FAILED TO USE A DENTAL DAM. THE STARTER DRILL FELL OUT OF THE EXTENDER AND THE PT SWALLOWED THE DRILL. THE PT WAS TAKEN TO THE EMERGENCY ROOM (ER) THE SAME DAY. THE ER PERSONNEL EXTRACTED THE STARTER DRILL AND THE PT WAS RELEASED THE SAME DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204771 BIOHORIZON DENTAL IMPLANT DRILL DENTAL IMPLANT DRILL DZA BIOHORIZONS 122-100 PO11031341A

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention