SYNCHROMED II
Report
- Report Number
- 3004209178-2013-07612
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO
- Product Code
- LKK
- PMA / PMN Number
- P860004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
UPON LABORATORY ARRIVAL AN X-RAY WAS TAKEN BEFORE THE RESERVOIR WAS ACCESSED. TEN MILLILITERS (ML) OF FLUID WERE ASPIRATED WHEN THE RESERVOIR WAS FIRST ACCESSED IN THE LABORATORY. THE PUMP PASSED ALL NON-DESTRUCTIVE TESTING, WITH RESERVOIR ACCESS ISSUES ENCOUNTERED. THE PUMP WAS THEN FILLED WITH 20 ML OF STERILE WATER, AND ASPIRATED A TOTAL OF FIVE TIMES, WITH NO PROBLEMS ENCOUNTERED. THE PUMP WAS THEN FILLED WITH 10 ML OF STERILE WATER, ASPIRATED A TOTAL OF FIVE TIMES, WITH NO PROBLEMS ENCOUNTERED. THE RESERVOIR ACCESS COMPLAINT COULD NOT BE DUPLICATED IN THE LABORATORY.
(B)(4).
ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THERE WERE DIFFICULTIES FILLING AND ASPIRATING THE PUMP. THE PHYSICIAN WAS ABLE, WITH DIFFICULTY, TO INJECT "ONLY 7 MLS" INTO THE PUMP. HE ALSO REPORTED DIFFICULTIES DRAINING THE PUMP. HE WAS ABLE TO REMOVE "ONLY 14 MLS." THE PUMP WAS NEVER IMPLANTED IN A PATIENT.
IT WAS LATER REPORTED THAT PRIOR TO THE PUMP IMPLANT PROCEDURE, THE SURGEON COULD NOT FILL THE PUMP. THE SURGEON EXPERIENCED ¿A LOT OF DIFFICULTIES TO EMPTY 4 OR 5 ML OF WATER.¿
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212833 | SYNCHROMED II | PUMP, INFUSION, IMPLANTED, PROGRAMMABLE | LKK | MDT PUERTO RICO OPERATIONS CO | 8637-20 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |