EON MINI
Report
- Report Number
- 1627487-2013-03612
- Event Type
- Injury
- Date Received
- May 8, 2013
- Date of Event
- April 17, 2013
- Report Date
- April 17, 2013
- Manufacturer
- ST. JUDE MEDICAL- NEUROMODULATION
- Product Code
- LGW
- Removal / Correction Number
- 1627487-07262012-002-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-03613. THE PATIENT REPORTED, SHE EXPERIENCES UNCOMFORTABLE HEATING AT HER SCS IPG POCKET SITE WHILE CHARGING AS WELL AS HER CHARGING SYSTEM PADDLE BECOMING HOT. THE PATIENT ALSO REPORTED THAT CHARGING TAKES TOO LONG. THE PATIENT HAS RECEIVED A REPLACEMENT CHARGING SYSTEM (LOW ENERGY). NO ADD¿L INFO IS AVAILABLE AT THIS TIME. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIV, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202448 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL- NEUROMODULATION | 3788 | 3422911 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention | SCS LEAD: MODEL 3186 (2)| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2) |