EON MINI
Report
- Report Number
- 1627487-2013-06263
- Event Type
- Injury
- Date Received
- May 8, 2013
- Date of Event
- March 25, 2013
- Report Date
- April 15, 2013
- Manufacturer
- ST. JUDE MEDICAL- NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-06264. IT WAS REPORTED THE PATIENT HAS STOPPED USING THE SCS SYSTEM BECAUSE THE CHARGER GETS HOT WHILE CHARGING. A NEW LOW ENERGY CHARGER WAS SENT TO ADDRESS THE ISSUE. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIV, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202365 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL- NEUROMODULATION | 3788 | 3750996 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention | SCS LEAD: MODEL 3228| IMPLANT DATE: |