CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-12639
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- November 15, 2012
- Report Date
- April 15, 2013
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS INCLUDED IN A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA). RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT #1627487-2013-12638. IT WAS REPORTED THE PT EXPERIENCED HEATING AND BURNS AT THE IPG POCKET SITE WHEN CHARGING. THE PT'S SKIN WAS PEELING AT THE BURN SITE. A NEW CHARGER WAS SENT TO THE PT. F/U REVEALED THE PT'S ISSUE HAS BEEN RESOLVED WITH THE NEW CHARGER. PT RETURNED THE OLD CHARGER TO SJM. ON (B)(4) 2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202517 | CHARGING SYSTEM | SCS CHARGER | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3721 | 3617312 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention | IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192| SCS ADAPTOR: MODEL 2342 (2) |