EON MINI
Report
- Report Number
- 1627487-2013-02641
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 18, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A 1627487-12192011-003-R. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION AND FIELD ADVISORIES. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-02642. IT WAS REPORTED THE PATIENT EXPERIENCES A CRAMPING PAIN RIGHT ABOVE WHERE HER IPG IS LOCATED. SHE ALSO REPORTED POCKET HEATING WHILE CHARGING. SHE STATED THE HEATING SENSATION AROUND HER IPG SITE BECOMES EXTREMELY UNCOMFORTABLE AND REQUIRES HER TO TAKE A BREAK FROM CHARGING. A NEW LE CHARGING SYSTEM WAS SENT TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204806 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3308526 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention | SCS LEAD, MODEL 3246 (2)| IMPLANT DATE: UNK |