EON MINI
Report
- Report Number
- 1627487-2013-02650
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 18, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REF MFR REPORT: 1627487-2013-02651. IT WAS REPORTED THE PT EXPERIENCED AN UNCOMFORTABLE HEATING SENSATION AT HER IPG SITE AFTER APPROX 10 MINUTES OF CHARGING. A NEW LE CHARGING SYSTEM WAS SENT TO THE PT. FOLLOW-UP INDICATED THE NEW CHARGER RESOLVED THE ISSUE. ON (B)(4) 2012, ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204780 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3730603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention | IMPLANT DATE:| IMPLANT DATE:| SCS LEAD, MODEL 3186 (2)| SCS ANCHOR, MODEL 1192 (2) |