FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3110636 · Received May 10, 2013

Report

Report Number
1627487-2013-08080
Event Type
Injury
Date Received
May 10, 2013
Date of Event
June 28, 2012
Report Date
April 16, 2013
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-08079. IT WAS REPORTED THE PT WAS EXPERIENCING AN UNCOMFORTABLE HEATING SENSATION WHILE CHARGING SINCE THE IMPLANT DATE. F/U INFO IDENTIFIED A REPLACEMENT CHARGER WAS SENT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
205282 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST JUDE MEDICAL - NEUROMODULATION 3721 3673730

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3186 (2)| SCS ANCHOR: MODEL 1192 (2)