CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-08080
- Event Type
- Injury
- Date Received
- May 10, 2013
- Date of Event
- June 28, 2012
- Report Date
- April 16, 2013
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-08079. IT WAS REPORTED THE PT WAS EXPERIENCING AN UNCOMFORTABLE HEATING SENSATION WHILE CHARGING SINCE THE IMPLANT DATE. F/U INFO IDENTIFIED A REPLACEMENT CHARGER WAS SENT TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 205282 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3721 | 3673730 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Required Intervention | IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3186 (2)| SCS ANCHOR: MODEL 1192 (2) |