EON MINI
Report
- Report Number
- 1627487-2013-08085
- Event Type
- Injury
- Date Received
- May 10, 2013
- Date of Event
- April 19, 2013
- Report Date
- April 19, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-08086. IT WAS REPORTED THE PT WAS EXPERIENCING HEATING WHILE CHARGING AND WAS ALSO HAVING SOME DISCOMFORT AT THE IPG SITE. A REPLACEMENT CHARGER WAS SENT TO THE PT AND THE NEW CHARGER RESOLVED THE ISSUE. ON (B)(4) 2012, ST. JUDE MEDICAL NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 207533 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3715870 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SCS LEAD: MODEL 3219| IMPLANT DATE: |