FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3110582 · Received May 14, 2013

Report

Report Number
2531779-2013-06320
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 17, 2013
Report Date
April 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

DEVICE EVALUATION - (B)(4): THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING RESULTS: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE 'OK','UP','DOWN' AND 'CONTRAST' BUTTONS. REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS. CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS. UNRELATED TO THIS COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED. T OPENING OF BATTERY CHAMBER EXTENDING DOWN TO THE PRIMARY SEAL. THE RETURNED BATTERY CAP WAS ABLE TO FULLY TIGHTEN UNTIL O-RING WAS SEATED AND CAP WAS FLUSH WITH THE PUMP CASE. NO POWER LOSS OR MOISTURE INGRESS WAS NOTED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS ARE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213121 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1