FDA Adverse Event
Malfunction
Summary report: N
GOBED II
MDR report key: 3110502
·
Received May 14, 2013
Report
- Report Number
- 0001831750-2013-04281
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- March 20, 2013
- Report Date
- March 20, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOWLER WAS DRIFTING DUE TO MALFUNCTION GAS CYLINDER. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212690 | GOBED II | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |