PRISM HBSAG CONFIRMATORY
Report
- Report Number
- 1415939-2013-00209
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- March 20, 2013
- Report Date
- April 25, 2013
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- KSJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
PATIENT IDENTIFIER : (B)(6). (B)(4). AN EVALUATION IS IN PROGRESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
PRODUCT EVALUATION WAS COMPLETED IN ORDER TO INVESTIGATE THIS ISSUE. EVALUATION OF COMPLAINT DATA FOR THE LOT IN QUESTION DID NOT IDENTIFY ATYPICAL ACTIVITY. REVIEW OF THE LOT DEVICE HISTORY RECORDS DID NOT REVEAL ANY ISSUES RELATED TO THIS INCIDENT. QUALITY RECORDS WERE REVIEWED TO DETERMINE IF OTHER CUSTOMERS HAD EXPERIENCED SIMILAR ISSUES THAT MIGHT WARRANT FURTHER INVESTIGATION. THE REVIEWS DID NOT SHOW ANY UNUSUAL ACTIVITY RELATED TO THIS ISSUE. BASED ON THE EVALUATION RESULTS, NO PRODUCT DEFICIENCY OR MALFUNCTION IDENTIFIED AND THE PRODUCT IS PERFORMING AS EXPECTED.
THE CUSTOMER QUESTIONED A (B)(6) RESULT FOR THE PRISM HBSAG CONFIRMATORY TEST WHEN TESTED A PATIENT SAMPLE (DONOR) FOR ANOTHER LAB WHERE THE DONOR SAMPLE HAD ALREADY BEEN TESTED REPEAT (B)(6) WITH THE PRISM HBSAG REAGENT. THE DONOR HAD ALWAYS BEEN (B)(6) FOR THIS ASSAY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213406 | PRISM HBSAG CONFIRMATORY | KSJ | ABBOTT LABORATORIES | 18045M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PRISM 6 CHANNEL-CE 06A36-04 SN (B)(4) |