FDA Adverse Event Malfunction Summary report: N

PRISM HBSAG CONFIRMATORY

MDR report key: 3110491 · Received May 14, 2013

Report

Report Number
1415939-2013-00209
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
March 20, 2013
Report Date
April 25, 2013
Manufacturer
ABBOTT LABORATORIES
Product Code
KSJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT IDENTIFIER : (B)(6). (B)(4). AN EVALUATION IS IN PROGRESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION WAS COMPLETED IN ORDER TO INVESTIGATE THIS ISSUE. EVALUATION OF COMPLAINT DATA FOR THE LOT IN QUESTION DID NOT IDENTIFY ATYPICAL ACTIVITY. REVIEW OF THE LOT DEVICE HISTORY RECORDS DID NOT REVEAL ANY ISSUES RELATED TO THIS INCIDENT. QUALITY RECORDS WERE REVIEWED TO DETERMINE IF OTHER CUSTOMERS HAD EXPERIENCED SIMILAR ISSUES THAT MIGHT WARRANT FURTHER INVESTIGATION. THE REVIEWS DID NOT SHOW ANY UNUSUAL ACTIVITY RELATED TO THIS ISSUE. BASED ON THE EVALUATION RESULTS, NO PRODUCT DEFICIENCY OR MALFUNCTION IDENTIFIED AND THE PRODUCT IS PERFORMING AS EXPECTED.

Description of Event or Problem · 1

THE CUSTOMER QUESTIONED A (B)(6) RESULT FOR THE PRISM HBSAG CONFIRMATORY TEST WHEN TESTED A PATIENT SAMPLE (DONOR) FOR ANOTHER LAB WHERE THE DONOR SAMPLE HAD ALREADY BEEN TESTED REPEAT (B)(6) WITH THE PRISM HBSAG REAGENT. THE DONOR HAD ALWAYS BEEN (B)(6) FOR THIS ASSAY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213406 PRISM HBSAG CONFIRMATORY KSJ ABBOTT LABORATORIES 18045M500

Patients

Seq Age Sex Outcome Treatment
1 PRISM 6 CHANNEL-CE 06A36-04 SN (B)(4)