FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3110272 · Received May 13, 2013

Report

Report Number
2531779-2013-06313
Event Type
Malfunction
Date Received
May 13, 2013
Report Date
April 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD IS FULLY INTACT. ALL KEYPAD BUTTONS RESPONDED APPROPRIATELY. THERE WAS NO DEFECT FOUND WITH ORIGINAL COMPLAINT. UNRELATED TO THE COMPLAINT, THE PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. PUMP COVER WAS REMOVED AND THE OLD SCREEN WAS REMOVED AND REPLACED WITH A NEW TEST SCREEN, CONTRAST RETURNED TO NORMAL WITH TEST SCREEN.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THE UP AND DOWN ARROW BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
210869 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 51 YR