FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 3110269 · Received May 13, 2013

Report

Report Number
9611451-2013-00380
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
April 15, 2013
Report Date
April 15, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RECEIVED AT FISHER & PAYKEL HEALTHCARE (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: INSPECTION OF THE COMPLAINT FEEDSET TUBE REVEALED THAT THE FEEDSET TUBE HAD BECOME SEPARATED FROM THE SPIKE. THERE WAS ADEQUATE GLUE ON THE FEEDSET TUBE/SPIKE CONNECTION BUT THE GLUE WAS NOT BONDING. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 121101. CONCLUSION: WE WERE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED FAULT. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE SPIKE BECAME LOOSE AFTER RELEASE FOR DISTRIBUTION, LIKELY AS A RESULT OF FAILURE OF THE GLUE BOND THAT JOINS THE SPIKE TO THE WATER FEEDSET TUBE. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT WATER LEAKAGE OCCURRED AT THE CONNECTION BETWEEN THE WATER FEED TUBE AND THE BAG SPIKE OF AN MR290 HUMIDIFICATION CHAMBER AT START OF USE. THIS OCCURRED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
211296 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1211010105

Patients

Seq Age Sex Outcome Treatment
1