ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-06296
- Event Type
- Malfunction
- Date Received
- May 13, 2013
- Report Date
- April 17, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW UP #1 SUBMITTED: 07/11/2013DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT. ON TESTING, ALL THE KEYPAD BUTTONS HAD INTERMITTENT RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE BUTTONS. THE DISPLAY WAS NOTED TO BE FADED, DISCOLORED AND DIFFICULT TO READ. A TEST DISPLAY WAS ATTACHED TO THE PUMP AND ILLUMINATED PROPERLY WITHOUT DISCOLORATION.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.