FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3110227
·
Received May 13, 2013
Report
- Report Number
- 3004209178-2013-93394
- Event Type
- Injury
- Date Received
- May 13, 2013
- Date of Event
- April 16, 2013
- Report Date
- April 17, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. HOWEVER, UNIT DID NOT COMMUNICATED PROPERLY WITH TEST DUE TO FAULTY COMPONENT ON THE DISPLAY BOARD.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT HE HAD UNEXPLAINED HIGH BLOOD GLUCOSE AND DKA. CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EMERGENCY ROOM WAS OVER 600 MG/DL. CUSTOMER STATED THAT HE HAD NO DELIVERY ALARMS PRIOR TO HOSPITALIZATION. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 210504 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization |