RESERVOIR 3ML
Report
- Report Number
- 3004209178-2013-93389
- Event Type
- Malfunction
- Date Received
- May 13, 2013
- Date of Event
- April 17, 2013
- Report Date
- April 17, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- PATIENT
Narratives
INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. INSPECTED THREE RANDOMLY SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THAT HE HAD HIGH BLOOD GLUCOSE DUE TO THE RESERVOIR WAS LEAKING INSULIN INTO RESERVOIR COMPARTMENT DURING NORMAL USE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 210067 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |