FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3110223 · Received May 13, 2013

Report

Report Number
3004209178-2013-93387
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
April 17, 2013
Report Date
April 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK AND CRACKED RESERVOIR TUBE NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE HAD HIGH BLOOD GLUCOSE AND THE INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
210460 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 40 YR