FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3110203 · Received May 13, 2013

Report

Report Number
3004209178-2013-93390
Event Type
Injury
Date Received
May 13, 2013
Date of Event
April 17, 2013
Report Date
April 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT WAS RECEIVED WITH INTERMITTENT BUTTON DUE TO MOISTURE DAMAGE AT KEYPAD TRACES. UNIT HAD CRACKED CASE AT DISPLAY WINDOW CORNERS AND SCRATCHED DISPLAY WINDOW. THE DRIVE SUPPORT WAS INSPECTED AND NO ANOMALY WAS NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD UNEXPLAINED HIGH BLOOD GLUCOSE, DKA AND WAS HOSPITALIZED IN (B)(6) OF 2012. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS UNKNOWN. CALLER STATED THAT THE CUSTOMER JUMPED INTO THE POOL WITH THE INSULIN PUMP. CALLER STATED THAT THE INSULIN PUMP KEYPAD BUTTONS ARE STICKING AND THE DRIVE SUPPORT CAP IS LOOSE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
211148 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 14 YR Hospitalization