FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3110050 · Received May 13, 2013

Report

Report Number
2032227-2013-01902
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
May 8, 2013
Report Date
May 8, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH LOOSE DRIVE SUPPORT DISK AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A LOOSE DRIVE SUPPORT CAP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
211235 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LWWB

Patients

Seq Age Sex Outcome Treatment
1