PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3110037 · Received May 13, 2013

Report

Report Number
2032227-2013-01920
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
April 30, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING NORMAL USE. THE CUSTOMER STATED THAT SHE HAD AN MRI SCAN RECENTLY, AND THE INSULIN PUMP WAS IN THE SAME ROOM, BUT WASN'T BEING WORN DURING THE SCAN. TROUBLESHOOTING WAS NOT POSSIBLE AS THE INSULIN PUMP WOULD NOT REWIND. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. THE REPORTED BLOOD GLUCOSE READING WAS 137 MG/DL. NOTHING FURTHER WAS REPORTED.