FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3110032 · Received May 13, 2013

Report

Report Number
2032227-2013-01915
Event Type
Malfunction
Date Received
May 13, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST, AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED AND SQUIRTED OUT INSULIN DURING THE MANUAL PRIME. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
211528 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LCAL

Patients

Seq Age Sex Outcome Treatment
1