SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART
Report
- Report Number
- 3003761017-2013-00054
- Event Type
- Injury
- Date Received
- May 6, 2013
- Date of Event
- April 18, 2013
- Report Date
- May 1, 2013
- Manufacturer
- SYNCARDIA SYSTEMS, INC.
- Product Code
- LOZ
- PMA / PMN Number
- P030011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DECISION WAS MADE TO REPLACE THE PATIENT'S TAH-T. AFTER BEING SUPPORTED BY TAH-T LOT 071778 FOR 839 DAYS, THE PATIENT'S TAH-T WAS REPLACED WITH TAH-T LOT 073904 ON (B)(6) 2013 WITHOUT INCIDENT. THE CUSTOMER REPORTED THAT AS OF (B)(6) 2013, THE PATIENT IS DOING WELL AND IS ABLE TO WALK AND SIT IN A CHAIR. ALTHOUGH HER LDH LEVELS ARE STILL HIGH, HER URINE IS CLEAR. AT THIS TIME, THERE IS NO INDICATION OF A MALFUNCTION OF THE EXPLANTED TAH-T, THE CSS CONSOLE OR THE FREEDOM DRIVER. THE EXPLANTED TAH-T WILL BE RETURNED TO SYNCARDIA FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WILL BE PROVIDED IN A SUPPLEMENTAL MDR.
THE PATIENT WAS IMPLANTED WITH THE SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T) ON (B)(6) 2010 AND IS CURRENTLY ENROLLED IN THE FREEDOM DRIVER SYSTEM IDE STUDY. THE CUSTOMER REPORTED THAT THE PATIENT STATED THAT SHE WAS NOT FEELING WELL. THE CUSTOMER ALSO REPORTED THAT THE FILL VOLUMES DISPLAYED ON THE FREEDOM DRIVER WERE IN THE 30S AND CARDIAC OUTPUT WAS IN THE UPPER 4 LITERS PER MINUTE RANGE. THE CUSTOMER ALSO REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL. ON (B)(6) 2013, SHE WAS SWITCHED FROM THE FREEDOM DRIVER TO A CSS CONSOLE AND HER CARDIAC OUTPUT IMPROVED, AND THE PATIENT STATED THAT SHE FELT BETTER. THE CUSTOMER ALSO REPORTED THAT DURING THE TIME WHEN THE PATIENT WAS BEING SUPPORTED BY THE CSS CONSOLE, SHE SHOWED SIGNS OF HEMOLYSIS AND BLOODY URINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 196038 | SYNCARDIA TEMPORARY TOTAL ARTIFICIAL HEART | ARTIFICIAL HEART | LOZ | SYNCARDIA SYSTEMS, INC. | TAH-T | 071778 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |