FDA Adverse Event Malfunction Summary report: N

DRILL BIT Ø1.8 L80/66 2FLUTE

MDR report key: 3105053 · Received May 10, 2013

Report

Report Number
1719045-2013-01424
Event Type
Malfunction
Date Received
May 10, 2013
Date of Event
April 8, 2013
Report Date
April 16, 2013
Manufacturer
SYNTHES MONUMENT
Product Code
HTW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. A DEVICE HISTORY REVIEW FOR THIS LOT HAS BEEN REQUESTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.

Additional Manufacturer Narrative · 1

THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. ORCHID UNIQUE MANUFACTURED THE 1.8 MM DRILL BIT/MINI QC-80 MM, PART 317.861, AND LOT NUMBER U161089. THE SUPPLIER CERTIFICATE OF COMPLIANCE, DATED NOVEMBER 8, 2012, INDICATES THE PARTS WERE MANUFACTURED TO PART 317.861, REVISION C AND MET THE REQUIRED SPECIFICATIONS. THE PARTS WERE INSPECTED AND CONFORMED TO THE SYNTHES, INCOMING FINAL INSPECTION SHEET NUMBER (B)(4), COMPLETED NOVEMBER 12, 2012. THERE WERE NO NON CONFORMITIES OR COMPLAINT RELATED ISSUES WITH THIS COMPLAINT. DUE TO AN UNKNOWN CAUSE, THE DRILL BIT APPEARED TO WOBBLE. THE MATERIAL LENGTH AND DIAMETER OF THE SHAFT WAS DETERMINED TO BE WITHIN SPECIFICATION.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE DISTRIBUTION STAFF CHECKED DRILL AND FOUND THEY WERE FLEXING ON (B)(6)2013. THE DRILL BIT WAS BENT. WHEN THEY ATTACHED THE DRILL WITH POWER TOOL, THE DRILL WAS WOBBLING. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
207096 DRILL BIT Ø1.8 L80/66 2FLUTE HTW SYNTHES MONUMENT U161089

Patients

Seq Age Sex Outcome Treatment
1