MEGA NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-01595
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- March 22, 2013
- Report Date
- April 18, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS PULLEY. THE IDLER PULLEY SPUN FREELY BUT HAS DAMAGES ON THE EDGE. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE MEGA NEEDLE DRIVER INSTRUMENT WAS NOTED TO HAVE A BROKEN CONNECTOR. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203881 | MEGA NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420194-08 | M10130102 920 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |