FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3103618 · Received May 9, 2013

Report

Report Number
2955842-2013-01590
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 3, 2013
Report Date
April 15, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT. ENGINEERING PLACED THE INSTRUMENT ON AN IN-HOUSE SYSTEM WITH THE TIP COVER, AND NO SMOKE AND BUBBLING WAS OBSERVED. THE TIP COVER WAS RETURNED WITH THE INSTRUMENT AND NO FITTING ISSUE WAS FOUND. ENGINEERING ALSO FOUND PAD PRINTING REMOVED AT MULTIPLE LOCATIONS ON THE TUBE EXTENSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT SMOKE AND BUBBLING WAS NOTED UNDER THE TIP COVER AT THE PROXIMAL END. THE TIP COVER WAS REMOVED AND REPLACED AND THE PROCEDURE WAS CONTINUED WITHOUT FURTHER OCCURRENCE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204520 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10 M10130211 317

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES