PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3103558
·
Received May 9, 2013
Report
- Report Number
- 3004209178-2013-93320
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- April 16, 2013
- Report Date
- April 16, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. THE DRIVE SUPPORT CAP WAS INSPECTED AND ANOMALY WAS NOTED. UNIT WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT HE WAS HOSPITALIZED DUE TO LOW BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS UNDER 30 MG/DL. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP IS PUSHED IN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203095 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Hospitalization |