FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3103558 · Received May 9, 2013

Report

Report Number
3004209178-2013-93320
Event Type
Injury
Date Received
May 9, 2013
Date of Event
April 16, 2013
Report Date
April 16, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. THE DRIVE SUPPORT CAP WAS INSPECTED AND ANOMALY WAS NOTED. UNIT WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE WAS HOSPITALIZED DUE TO LOW BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS UNDER 30 MG/DL. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP IS PUSHED IN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203095 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 31 YR Hospitalization