COULTER® LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-00772
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K032000
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE ON (B)(4) 2013. THE FSE INSPECTED THE INSTRUMENT AND WAS ABLE TO CONFIRM A LEAK AT THE NEEDLE LOCATION. THE FSE DISCOVERED THAT THE NEEDLE BELLOWS WAS NOT DRAINING PROPERLY AND CAUSING BLOODY DILUENT TO OVERFLOW OUT OF THE TOP OF THE NEEDLE BELLOWS ASSEMBLY. THE FSE PERFORMED A BLEACH PROCEDURE ON THE NEEDLE BELLOWS DRAIN PATHWAY AND REMOVED A CLOT FROM THE PATHWAY. FAILURE MODE WAS A CLOT IN THE NEEDLE BELLOWS ASPIRATION/DRAIN PATHWAY. HOWEVER, THE INSTRUMENT PROVIDED ASPIRATION ERRORS AND/OR NO CONTROL RESULTS TO ALERT THE OPERATOR TO AN INSTRUMENT PROBLEM. (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) AND STATED THAT THERE WAS A BLOODY FLUID LEAK (1CC) UNDER THE ROCKERBED NEAR THE NEEDLE CARTRIDGE, WHEN PERFORMING MORNING START-UP ON THE COULTER LH 500 HEMATOLOGY ANALYZER. PARTIAL ASPIRATIONS ERRORS AND NO RESULTS WERE OCCURRING ON CONTROLS. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THERE IS AN EXPOSURE CONTROL PLAN AT THE FACILITY. THE OPERATOR WAS WEARING A LABORATORY COAT, GLOVES AND FACE PROTECTION. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. NO PATIENT RESULTS WERE RUN AS THE CUSTOMER COULD NOT GET CONTROLS TO RUN. NO PATIENT RESULTS WERE AFFECTED. THERE WAS NO DEATH, INJURY OR EFFECT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203302 | COULTER® LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 500 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |