FDA Adverse Event Malfunction Summary report: N

MULTI-LINK MINI VISION RX CORONARY STENT SYSTEM

MDR report key: 3103394 · Received May 9, 2013

Report

Report Number
2024168-2013-02922
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 8, 2013
Report Date
April 17, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
P020047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT DETACHMENT WAS CONFIRMED. THE REPORTED FAILURE TO CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE HEAVILY CALCIFIED CIRCUMFLEX. DURING THE FAILED ATTEMPT TO CROSS THE LESION WITH THE 2.25X23 MM MINI VISION STENT DELIVERY SYSTEM, THE PROXIMAL SHAFT SEPARATED. A NON-ABBOTT STENT WAS USED PRIOR TO THE MINI VISION AND IT ALSO FAILED TO CROSS THE LESION. NO DEVICE WAS SUCCESSFUL IN CROSSING THE LESION. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204297 MULTI-LINK MINI VISION RX CORONARY STENT SYSTEM CORONARY STENT SYSTEM MAF AV-TEMECULA-CT 1070641

Patients

Seq Age Sex Outcome Treatment
1 79 YR