MULTI-LINK MINI VISION RX CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2013-02922
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 17, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MAF
- PMA / PMN Number
- P020047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT DETACHMENT WAS CONFIRMED. THE REPORTED FAILURE TO CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE HEAVILY CALCIFIED CIRCUMFLEX. DURING THE FAILED ATTEMPT TO CROSS THE LESION WITH THE 2.25X23 MM MINI VISION STENT DELIVERY SYSTEM, THE PROXIMAL SHAFT SEPARATED. A NON-ABBOTT STENT WAS USED PRIOR TO THE MINI VISION AND IT ALSO FAILED TO CROSS THE LESION. NO DEVICE WAS SUCCESSFUL IN CROSSING THE LESION. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204297 | MULTI-LINK MINI VISION RX CORONARY STENT SYSTEM | CORONARY STENT SYSTEM | MAF | AV-TEMECULA-CT | 1070641 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |