ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06117
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Report Date
- April 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS:BLACK BOX INDICATES A TIME AND DATE RESET IN WITHIN 10 MINUTES OF POWER OFF. TIME AND DATE WAS ACCURATELY SET AT START OF INVESTIGATION. PERFORMED PUMP REWIND, LOAD, AND PRIME WITH NO ALARMS. PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE. THE BATTERY WAS REMOVED FOR ABOUT 6 HOUR AND REINSERTED, THE PUMP TIME AND DATE DID RESET TO THE DEFAULT SETTINGS. PUMP WAS OPEN TO INVESTIGATE, COMPONENT BT1 WAS FOUND TO BE LEAKING. UNRELATED TO THIS COMPLAINT, PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. PUMP COVER WAS REMOVED AND THE OLED SCREEN WAS REMOVED AND REPLACED WITH A NEW TEST SCREEN, CONTRAST RETURNED TO NORMAL WITH TEST SCREEN. THERE WAS A VISIBLE CRACK IN THE BATTERY COMPARTMENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED DIM/DISCOLORED DISPLAY, CRACKED BATTERY COMPARTMENT, AND A LEAKING INTERNAL BATTERY. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203129 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |