ACUITY
Report
- Report Number
- 2124215-2013-05876
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- March 19, 2013
- Report Date
- May 6, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- PMA / PMN Number
- G050163
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. THE LID TIP HAS NO VISIBLE SIGNS OF DAMAGE OR DEFECT WHICH WOULD LEAD TO DISLODGMENT. THE ANALYSIS RESULT COULD NOT CONFIRM THE ALLEGATION OF DISLODGMENT.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD HAD DISLODGED. THE LV LEAD WAS EXPLANTED AND NO LONGER IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203679 | ACUITY | IMPLANTABLE HF LEAD | OJX | CPI - DEL CARIBE | 4592 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| L| R | 4470| 4471| 4554| 5071| MISMATCH| 4592| 0276| N140| 1298 |