D-M 1.6MM BEADED CABLE SET VIT
Report
- Report Number
- 0002249697-2013-01601
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 16, 2013
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JDQ
- PMA / PMN Number
- K953818
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
THERE WERE NO REPORTED DISCREPANCIES AND THERE WERE NO OTHER EVENTS FOR THE REFERENCED LOT. THE RETURNED CABLE AND SLEEVE ARE IN PRISTINE CONDITION. VISUAL INSPECTION FOUND NO DISCREPANCIES. FUNCTIONAL INSPECTION FOUND THE CABLE CAN BE FULLY PASSED THROUGH BOTH HOLES OF THE SLEEVE WITH NO ISSUE. INSUFFICIENT MEDICAL RECORDS INFORMATION WAS RECEIVED FOR REVIEW WITH A CLINICIAN CONSULTANT. THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE THE DEVICE PASSED THE FUNCTIONAL INSPECTION. FURTHER INFORMATION SUCH AS RETURN OF TENSIONER DEVICE, OPERATIVE REPORT AS WELL AS PATIENT HISTORY AND FOLLOW-UP NOTES ARE NEEDED TO COMPLETE THE INVESTIGATION FOR DETERMINING ROOT CAUSE. THE REPORTED EVENT REGARDING ASSEMBLY ISSUE INVOLVING A DALL MILE CABLE SET WAS NOT CONFIRMED.
IT WAS REPORTED THAT THE CABLE COULD NOT BE INSERTED INTO THE SINGLE SIDE TENSIONER. SO, THE SURGEON USED SPARE CABLE INTO THE SAME SINGLE SIDE TENSIONER,AND IT COULD BE INSERTED.
IT WAS REPORTED THAT THE CABLE COULD NOT BE INSERTED INTO THE SINGLE SIDE TENSIONER. SO, THE SURGEON USED SPARE CABLE INTO THE SAME SINGLE SIDE TENSIONER, AND IT COULD BE INSERTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203390 | D-M 1.6MM BEADED CABLE SET VIT | IMPLANT | JDQ | STRYKER ORTHOPAEDICS-MAHWAH | 41046502 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Other |