FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ?

MDR report key: 3103263 · Received May 9, 2013

Report

Report Number
2134265-2013-03021
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 11, 2013
Report Date
April 12, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: THE COMPLAINT DEVICE WAS RECEIVED FOR ANALYSIS. A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE NOTED PROXIMAL STENT DAMAGE. STRUTS AT THE PROXIMAL EDGE OF THE STENT WERE BUNCHED AND MISALIGNED. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS.  THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, STENT DAMAGE OCCURRED. THE 90% STENOSED AND SEVERELY CALCIFIED LESION WAS LOCATED IN THE SEVERELY TORTUOUS MIDDLE OF LEFT ANTERIOR DESCENDING ARTERY. A 3.00X20MM PROMUS ELEMENT DRUG-ELUTING STENT WAS USED TO CROSS THE LESION, BUT NO CROSS OCCURRED. WHEN THE PHYSICIAN REMOVED THE DEVICE, THEY NOTICED THAT A PART OF THE STENT WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, STENT DAMAGE OCCURRED. THE 90% STENOSED AND SEVERELY CALCIFIED LESION WAS LOCATED IN THE SEVERELY TORTUOUS MIDDLE OF LEFT ANTERIOR DESCENDING ARTERY. A 3.00X20MM PROMUS ELEMENT DRUG-ELUTING STENT WAS USED TO CROSS THE LESION, BUT NO CROSS OCCURRED. WHEN THE PHYSICIAN REMOVED THE DEVICE, THEY NOTICED THAT A PART OF THE STENT WAS LIFTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202876 PROMUS ELEMENT ? STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911320300 15399286

Patients

Seq Age Sex Outcome Treatment
1